CA 8:1 Endo

Bien-Air CA 8:1 Endo contra-angle is designed to work with shorter endo file shanks. Offering unparalleled visibility in molar areas, its small head provides the ideal solution for rotary endodontic procedures.

Reference:

CA 8:1 Endo 1600911-001

CA 8:1 Endo
 

Technical Data

CA 8:1 Endo
Transmission ratio8:1
Maximum speed (rpm)5’000
Head diameter (mm)8.5
Head size (mm)9.4
Weight (g)46
Coupling systemISO 3694 and Bien-Air Micro-Series
Warranty1 to 2 years 1
Reference CA 8:1 Endo 1600911-001
CA 8:1 Endo
Transmission ratio 8:1
Maximum speed (rpm) 5’000
Head diameter (mm) 8.5
Head size (mm) 9.4
Weight (g) 46
Coupling system ISO 3694 and Bien-Air Micro-Series
Warranty 1 to 2 years 1
Reference CA 8:1 Endo 1600911-001

1 1-year standard warranty and 1-year optional warranty available through Bien-Air’s PlanCare extended warranty program.

Maintenance

Maintenance of contra-angles

PREPARATION
A.

REMOVING THE INSTRUMENT

  • Remove the rotary instrument from the chuck mechanism
  • Disconnect the contra-angle from the motor
B.

PRE-DISINFECTION

  • External disinfection

Do not immerse the instrument in a disinfectant bath.

CLEANING
A.

CLEAN UNDER RUNNING WATER AT <38°C

B.

REMOVING DIRT/DEPOSITS

  • Carefully remove all traces of dirt or deposits from the nozzles using the Bien-Air cleaning wire
  • Dry the nozzles with compressed air then dry the rest of the instrument with a cloth

Do not immerse in an ultrasonic cleaner

C.

INTERNAL AND EXTERNAL CLEANING WITH SPRAYNET

  • Shake the Spraynet before use
  • Use the Spraynet attachment 1
  • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Spraynet button down for 1 second
  • Clean the exterior of the contra-angle with a cloth soaked in Spraynet
DISINFECTION
A.

DISINFECTION USING SURFACE DISINFECTANT (follow the manufacturer's instructions for use)

Do not immerse the instrument in a disinfectant bath

OR
B.

THERMAL DISINFECTION (follow the manufacturer's instructions for use)

LUBRICATION
A.

LUBRICATION WITH LUBRIFLUID

  • Shake the Lubrifluid before usen
  • Use the Lubrifluid attachment 1
  • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Lubrifluid button down for 1 second
  • Then, run an operating test for 30 seconds at low speed to remove any surplus lubricant
  • Store immediately: store the instruments with the tip pointing downwards so that any oil residues can flow out
STERILISATION
A.

PLACE THE CONTRA-ANGLE IN A STERILISATION BAG

B.

STERILISE IN ACCORDANCE WITH THE MANUFACTURER'S INSTRUCTIONS

  • Instruments bearing the "135°C" can be sterilised in an autoclave

Only sterilise instruments which have been cleaned and lubricated