CA

CA 1:2.5 Versatile and ergonomic

The new CA 1:2.5 L Micro-Series is the gold standard contra-angle for surgery and periodontics.

Thanks to its speed increasing ratio, the CA 1:2.5 L Micro-Series can be used for a wide range of clinical applications : apicectomy,  crown removal and crown lengthening.

Reference:

CA 1:2.5 CA 1:2.5

With the irrigation line routed inside the CA 1:2.5 L Micro-Series, you have complete freedom of movement, with nothing to obstruct your handling of the instrument.

Connected to the MX-i PLUS micromotor, the CA 1:2.5 L Micro-Series offers you perfect balance and unrivalled ease of use.

Multi-strand optical
glass technology


Bench test for durability: over 1500 cycles at 70 Ncm

Technical Data

CA 1:2.5 L Micro-Series
Transmission ratio1:2.5
Maximum speed (rpm)100’000
Head-
Head diameter (mm)9.8
Head size (mm)15.5 (22.9 mm with bur 19 mm)
Weight (g)82
Noise level (dBa)57
Spray system2 x 3 separate air/water sprays
Lighting systemMulti-strand optical glass technology
CouplingISO 3964 & Bien-Air Micro-Series
Warranty1  year 
Reference CA 1:2.5 L Micro-Series 1601055-001
CA 1:2.5 L Micro-Series
Transmission ratio 1:2.5
Maximum speed (rpm) 100’000
Head -
Head diameter (mm) 9.8
Head size (mm) 15.5 (22.9 mm with bur 19 mm)
Weight (g) 82
Noise level (dBa) 57
Spray system 2 x 3 separate air/water sprays
Lighting system Multi-strand optical glass technology
Coupling ISO 3964 & Bien-Air Micro-Series
Warranty 1  year 
Reference CA 1:2.5 L Micro-Series 1601055-001

Maintenance

MAINTENANCE OF CONTRA-ANGLES AND HANDPIECES FOR SURGERY AND IMPLANTOLOGY

PREPARATION
A.

REMOVING THE INSTRUMENT

  • Remove the rotary instrument from the chuck mechanism
  • Disconnect the contra-angle from the motor
B.

PRE-DISINFECTION

  • External disinfection

Do not immerse the instrument in a disinfectant bath.

CLEANING
A.

CLEAN UNDER RUNNING WATER AT <38°C

B.

REMOVING DIRT/DEPOSITS

  • Carefully remove all traces of dirt or deposits from the nozzles using the Bien-Air cleaning wire
  • Dry the nozzles with compressed air then dry the rest of the instrument with a cloth

Do not immerse in an ultrasonic cleaner

C.

INTERNAL CLEANING WITH AQUACARE

  • Shake the Aquacare before use
  • Use the Aquacare attachment 1
  • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Aquacare button down for 2 seconds
  • Use the Aquacare attachment 3
  • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Aquacare button down for 1 second
D.

INTERNAL AND EXTERNAL CLEANING WITH SPRAYNET

  • Shake the Spraynet before use
  • Use the Spraynet attachment
  • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Spraynet button down for 1 second
  • Clean the exterior of the contra-angle with a cloth soaked in Spraynet
DISINFECTION
A.

DISINFECTION USING SURFACE DISINFECTANT (follow the manufacturer's instructions for use)

Do not immerse the instrument in a disinfectant bath

OR
B.

THERMAL DISINFECTION (follow the manufacturer's instructions for use)

LUBRICATION
A.

LUBRICATION WITH LUBRIFLUID

  • Shake the Lubrifluid before use
  • Use the Lubrifluid attachment 1
  • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Lubrifluid button down for 1 second
  • Then, run an operating test for 30 seconds at a low speed to remove any surplus lubricant
  • Store immediately: store the instruments with the tip pointing downwards so that any oil residues can flow out
STERILISATION
A.

PLACE THE CONTRA-ANGLE IN A STERILISATION BAG

B.

STERILISE IN ACCORDANCE WITH THE MANUFACTURER'S INSTRUCTIONS

  • Instruments bearing the "135°C" symbol can be sterilised in an autoclave

Only sterilise instruments which have been cleaned and lubricated