Prophy

Handpieces with interchangeable heads and contra-angles with detachable cups are the only instruments that can be fitted to motors rotating at 40,000 rpm. Apart from superior comfort and extended service life,Bien-Air prophylaxis instruments in particular offer high resistance to brasive cleaning pastes and can withstand constant and repeated sterilizations.

Reference:

CAP 15:1 Prophy 1600290-001

PMP 10:1 PROPHY 1600289-001

Technical Data

CAP 15:1 Prophy PMP 10:1 Prophy Straight
TypeContra-angle without light or spray, screw or clip cup interchangeabilityStraight handpiece for standard disposable heads.
Ratio15:1 (speed reducing factor of 15)10:1 (speed reducing factor of 10)
Cleaning / sterilisationThermal washer / steam autoclave up to 134°C (273°F)Thermal washer / steam autoclave up to 134°C (273°F)
Warranty2 to 3 years 12 to 3 years 1
Reference CAP 15:1 Prophy 1600290-001 PMP 10:1 Prophy Straight 1600289-001
CAP 15:1 Prophy
Type Contra-angle without light or spray, screw or clip cup interchangeability
Ratio 15:1 (speed reducing factor of 15)
Cleaning / sterilisation Thermal washer / steam autoclave up to 134°C (273°F)
Warranty 2 to 3 years 1
Reference CAP 15:1 Prophy 1600290-001
PMP 10:1 Prophy Straight
Type Straight handpiece for standard disposable heads.
Ratio 10:1 (speed reducing factor of 10)
Cleaning / sterilisation Thermal washer / steam autoclave up to 134°C (273°F)
Warranty 2 to 3 years 1
Reference PMP 10:1 Prophy Straight 1600289-001

1 2-year standard warranty and 1-year optional warranty available through Bien-Air’s PlanCare extended warranty program.

Maintenance

Maintenance of contra-angles

PREPARATION
A.

REMOVING THE INSTRUMENT

  • Remove the rotary instrument from the chuck mechanism
  • Disconnect the contra-angle from the motor
B.

PRE-DISINFECTION

  • External disinfection

Do not immerse the instrument in a disinfectant bath.

CLEANING
A.

CLEAN UNDER RUNNING WATER AT <38°C

B.

REMOVING DIRT/DEPOSITS

  • Carefully remove all traces of dirt or deposits from the nozzles using the Bien-Air cleaning wire
  • Dry the nozzles with compressed air then dry the rest of the instrument with a cloth

Do not immerse in an ultrasonic cleaner

C.

INTERNAL AND EXTERNAL CLEANING WITH SPRAYNET

  • Shake the Spraynet before use
  • Use the Spraynet attachment 1
  • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Spraynet button down for 1 second
  • Clean the exterior of the contra-angle with a cloth soaked in Spraynet
DISINFECTION
A.

DISINFECTION USING SURFACE DISINFECTANT (follow the manufacturer's instructions for use)

Do not immerse the instrument in a disinfectant bath

OR
B.

THERMAL DISINFECTION (follow the manufacturer's instructions for use)

LUBRICATION
A.

LUBRICATION WITH LUBRIFLUID

  • Shake the Lubrifluid before usen
  • Use the Lubrifluid attachment 1
  • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Lubrifluid button down for 1 second
  • Store immediately: store the instruments with the tip pointing downwards so that any oil residues can flow out
STERILISATION
A.

PLACE THE CONTRA-ANGLE IN A STERILISATION BAG

B.

STERILISE IN ACCORDANCE WITH THE MANUFACTURER'S INSTRUCTIONS

  • Instruments bearing the "135°C" can be sterilised in an autoclave

Only sterilise instruments which have been cleaned and lubricated