Prophy
Handpieces with interchangeable heads and contra-angles with detachable cups are the only instruments that can be fitted to motors rotating at 40,000 rpm. Apart from superior comfort and extended service life,Bien-Air prophylaxis instruments in particular offer high resistance to brasive cleaning pastes and can withstand constant and repeated sterilizations.
Reference:
CAP 15:1 Prophy 1600290-001
PMP 10:1 PROPHY 1600289-001
Technical Data
CAP 15:1 Prophy
PMP 10:1 Prophy Straight
Type Contra-angle without light or spray, screw or clip cup interchangeability Straight handpiece for standard disposable heads. Ratio 15:1 (speed reducing factor of 15) 10:1 (speed reducing factor of 10) Cleaning / sterilisation Thermal washer / steam autoclave up to 134°C (273°F) Thermal washer / steam autoclave up to 134°C (273°F) Warranty 2 to 3 years 1 2 to 3 years 1
Reference
CAP 15:1 Prophy 1600290-001
PMP 10:1 Prophy Straight 1600289-001
CAP 15:1 Prophy
Type
Contra-angle without light or spray, screw or clip cup interchangeability
Ratio
15:1 (speed reducing factor of 15)
Cleaning / sterilisation
Thermal washer / steam autoclave up to 134°C (273°F)
Warranty
2 to 3 years 1
Reference
CAP 15:1 Prophy 1600290-001
PMP 10:1 Prophy Straight
Type
Straight handpiece for standard disposable heads.
Ratio
10:1 (speed reducing factor of 10)
Cleaning / sterilisation
Thermal washer / steam autoclave up to 134°C (273°F)
Warranty
2 to 3 years 1
Reference
PMP 10:1 Prophy Straight 1600289-001
Maintenance
Maintenance of contra-angles
PREPARATION
A.
REMOVING THE INSTRUMENT
Remove the rotary instrument from the chuck mechanism
Disconnect the contra-angle from the motor
B.
PRE-DISINFECTION
Do not immerse the instrument in a disinfectant bath.
CLEANING
A.
CLEAN UNDER RUNNING WATER AT <38°C
B.
REMOVING DIRT/DEPOSITS
Carefully remove all traces of dirt or deposits from the nozzles using the Bien-Air cleaning wire
Dry the nozzles with compressed air then dry the rest of the instrument with a cloth
Do not immerse in an ultrasonic cleaner
C.
INTERNAL AND EXTERNAL CLEANING WITH SPRAYNET
Shake the Spraynet before use
Use the Spraynet attachment 1
Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Spraynet button down for 1 second
Clean the exterior of the contra-angle with a cloth soaked in Spraynet
DISINFECTION
A.
DISINFECTION USING SURFACE DISINFECTANT (follow the manufacturer's instructions for use)
Do not immerse the instrument in a disinfectant bath
OR
B.
THERMAL DISINFECTION (follow the manufacturer's instructions for use)
LUBRICATION
A.
LUBRICATION WITH LUBRIFLUID
Shake the Lubrifluid before usen
Use the Lubrifluid attachment 1
Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Lubrifluid button down for 1 second
Store immediately: store the instruments with the tip pointing downwards so that any oil residues can flow out
STERILISATION
A.
PLACE THE CONTRA-ANGLE IN A STERILISATION BAG
B.
STERILISE IN ACCORDANCE WITH THE MANUFACTURER'S INSTRUCTIONS
Instruments bearing the "135°C" can be sterilised in an autoclave
Only sterilise instruments which have been cleaned and lubricated