Maintenance of Turbines
PREPARATION
A.
REMOVING THE BUR
- Remove the rotary bur from the chuck mechanism.
PRELIMINARY CLEANING
A.
RINSING THE DEVICE
- Clean the exterior and the interior of the device under tap water
B.
REMOVING DIRT/DEPOSITS
- Dry the nozzles then the rest of the instrument with
a sterile non-woven compress (low linting textiles).
DISINFECTION
A.
CLEANING AND DISINFECTION
- Dip the device in a bath containing
a cleaning product/disinfectant
- Brush the device with a smooth,
flexible brush. DO NOT USE a wire brush.
- Rinse the device twice under running tap water.
B.
OPTIONAL: AUTOMATIC DISINFECTION
- Carry out automatic washing or cleaning/disinfection.**
- After cleaning/disinfection, check that there are
no detergent residues and, if required, rinse again.
- After cleaning/disinfection, check that the device
is dry. If required, manually dry the device (see 4).
**For more information, please refer
to the manufacturer instructions for use
MANUAL CLEANING AND DRYING
A.
INTERNAL AND EXTERNAL CLEANING AND DRYING WITH SPRAYNET
- Shake the Spraynet before use.
- Select the appropriate Spraynet 1 nozzle*
- Keep the tip of the instrument pointing down
and protect it with a sterile non-woven cloth
- Activate the Spraynet.
- Dry the exterior of the turbine
with a cloth soaked in Spraynet.
Always dry the device.
Lubrification
A.
LUBRICATION WITH LUBRIMED
(For 2-/3-way and 4-way turbines, see inset below)
ou
B.
LUBRICATION WITH LUBRIFLUID
- Shake the Lubrifluid before use.
- Use the Lubrifluid attachment 1*
- Select the appropriate Lubrifluid nozzle 2*
- Keep the tip of the instrument pointing
down and protect it with an absorbent cloth.
- Activate the Lubrifluid
FOR 2-/3-WAY AND 4-WAY TURBINES ONLY
B'.
INTERNAL LUBRICATION WITH LUBRIFLUID
FOR 2-/3-WAY AND 4-WAY TURBINES ONLY
- Use the Lubrifluid attachment 1
- Select the appropriate Lubrifluid nozzle 3*
- Keep the tip of the instrument pointing down
and protect with an absorbent cloth; press
the Lubrifluid button down for 1 second.
Spray into the opening as shown in the
illustrations for “2-/3-way” and “4-way”.
STERILIZATION
A.
PLACE THE TURBINE IN A STERILIZATION BAG
B.
STERILIZE IN ACCORDANCE WITH
THE MANUFACTURER‘S INSTRUCTIONS
- Only instruments bearing the "135°C" symbol
can be sterilized in an autoclave.
Only sterilize instruments which have been
properly cleaned, dried and lubricated.
Download the maintenance instructions
Other maintenance instructions
Oral Surgery Implantology Handpieces
Our maintenance product range