Maintenance of Oral Surgery Implantology Handpieces
PREPARATION
A.
REMOVING THE INSTRUMENT
- Remove the rotary instrument from the chuck mechanism
- Disconnect the contra-angle from the motor
B.
PRE-DISINFECTION / ONLY FOR CA 1:2.5
Do not immerse the instrument in a disinfectant bath.
CLEANING
A.
CLEAN UNDER RUNNING WATER AT <38°C
B.
REMOVING DIRT/DEPOSITS / ONLY FOR CA 1:2.5
- Carefully remove all traces of dirt or deposits from the nozzles using the Bien-Air cleaning wire
- Dry the nozzles with compressed air then dry the rest of the instrument with a cloth
Do not immerse in an ultrasonic cleaner
C.
INTERNAL CLEANING WITH AQUACARE
- Shake the Aquacare before use
- Use the Aquacare attachment 1
- Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Aquacare button down for 2 seconds
- Use the Aquacare attachment 3
- Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Aquacare button down for 1 second
D.
INTERNAL AND EXTERNAL CLEANING WITH SPRAYNET
- Shake the Spraynet before use
- Use the Spraynet attachment
- Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Spraynet button down for 1 second
- Clean the exterior of the contra-angle with a cloth soaked in Spraynet
DISINFECTION
A.
DISINFECTION USING SURFACE DISINFECTANT (follow the manufacturer's instructions for use)
Do not immerse the instrument in a disinfectant bath
OR
B.
THERMAL DISINFECTION (follow the manufacturer's instructions for use)
LUBRICATION
A.
LUBRICATION WITH LUBRIFLUID
- Shake the Lubrifluid before use
- Use the Lubrifluid attachment 1
- Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Lubrifluid button down for 1 second
- Then, run an operating test for 30 seconds at a low speed to remove any surplus lubricant
- Store immediately: store the instruments with the tip pointing downwards so that any oil residues can flow out
STERILISATION
A.
PLACE THE CONTRA-ANGLE IN A STERILISATION BAG
B.
STERILISE IN ACCORDANCE WITH THE MANUFACTURER'S INSTRUCTIONS
- Instruments bearing the "135°C" symbol can be sterilised in an autoclave
Only sterilise instruments which have been cleaned and lubricated